AB200 |
pH Mode |
Range |
-2.000 to 20.000 |
Resolution |
0.1 / 0.01 / 0.001 |
Accuracy |
±0.002 pH + 1 LSD |
Buffer Sets |
USA: 2.000, 4.010, 6.997, 10.013, 12.000
NIST: 1.678, 4.010, 6.865, 9.184, 12.460
DIN: 1.090, 3.060, 4.650, 6.790, 9.230, 12.750
FSCI: 1.000, 3.000, 6.000, 8.000, 10.000, 13.000
CUSTOM: Any 2-5 values, ≥1.0 pH unit apart |
Slope Display |
Yes, up to 5 different slopes with offset
|
Temp Compensation |
Automatic or manual (0 to 100 ºC / 32 to 212 ºF)
|
mV Mode |
Range |
±2000 mV / Rel mV |
Resolution |
0.1mV |
Accuracy |
±0.2 mV or ±0.05 % whichever is greater |
Offset Adjustment |
Up to ±150 mV |
Temperature Mode |
Resolution |
0.001 to 19999 (±2000 mV) |
Accuracy |
0.001 / 0.01 / 0.1 / 1 (automatic) |
Calibration |
ppm, mg/L, molar |
Conductivity Mode |
Range |
0.00 μS to 500.0 mS |
Resolution |
0.01 / 0.1 μS ; 0.001 / 0.01 / 0.1 mS |
Accuracy |
±1 % full scale |
Coefficient (Per ºC) |
Linear & pure ; adjustable 0.000 to 10.000 % |
Normalization |
15.0 to 30.0 ºC |
TDS Mode |
Range |
0.00 ppm to 500 ppt (@ TDS factor 1.00) |
Resolution |
0.01 / 0.1 ppm ; 0.001 / 0.01 / 0.1 ppt |
Accuracy |
±1 % full scale |
TDS Factor |
0.400 to 1.000 |
Salinity Mode |
Range |
0 to 80.0 ppt |
Resolution |
0.01 / 0.1 ppm ; 0.001 / 0.01 / 0.1 ppt |
Accuracy |
±1 % full scale |
Resistivity Mode |
Range |
2.000 Ω to 20.0 MΩ |
Resolution |
0.01 / 0.1 Ω ; 0.001 / 0.1 kΩ ; 0.01 MΩ |
Accuracy |
±1 % full scale |
Output |
RS-232 (phono plug), mini-B USB, stirrer |
Language Selection |
English, Deutsch, 中文, Français, Italiano, & Español |
Memory |
500 data sets, viewable |
Datalogging |
Manual, timed (selectable every 3 to 3600 seconds)Printer or CSV format |
Cal Due Alarm |
Yes, user selectable from 8 hrs, 16 hrs, & 1-31 days |
High / Low Alarms |
Yes, user selectable, visual & audible |