| Compared products |
AS 120.R1 PLUS Analytical Balance |
AS 220.R1 PLUS Analytical Balance |
| Metrological parameters |
| Maximum capacity [Max] |
120 g |
220 g |
| Minimum load |
10 mg |
10 mg |
| Readability [d] |
0,1 µg |
0,1 µg |
| Tare range |
-120 g |
-220 g |
| Standard repeatability [5% Max] |
0,6 µg |
0,6 µg
|
| Standard repeatability [Max] |
0,06 mg |
0,07 mg |
| Standard minimum weight (USP) |
120 mg |
140 mg |
| Standard minimum weight (U=1%, k=2) |
12 mg |
14 mg |
| Permissible repeatability [5% Max] |
0,09 mg |
0,09 mg |
| Permissible repeatability [Max] |
0,1 mg |
0,1 mg |
| Linearity |
±2 µg |
±2 µg |
| Stabilization time |
2 s |
2 s |
| Adjustment |
external |
external |
| OIML Class |
- |
- |
| Physical parameters |
| Leveling system |
manual |
| Display |
LCD (backlit) |
| Protection class |
IP 43 |
| Delivery components |
Balance, weighing pan, weighing pan shield, centring ring, bottom cover, power supply. |
Balance, weighing pan, weighing pan shield, bottom cover, power supply. |
| Weighing pan dimensions |
Ø 100 mm |
| Packaging dimensions |
495×400×515 mm |
490×400×520 mm |
| Net weight |
7,3 kg |
| Gross weight |
9,3 kg |
| Communication interface |
| Communication interface |
2×RS232, 2×USB-A (interchangeable), USB-B, Wi-Fi (option) |
| Electrical parameters |
| Power supply |
Adapter: 100 – 240V AC 50/60Hz 0,6A; 12V DC 1,2A Balance: 12 – 15V DC 0,4A max |
| Environmental conditions |
| Operating temperature |
+10 ÷ +40 °C |