Specifications and Ordering Information |
4 ft models |
6 ft models |
Standard Model) |
Coated interior walls, cabinet only |
1354 |
1359 |
Standard Package |
Includes coated interior cabinet walls, manual
adjustable height stand, factory installed UV light,
and one set of armrests |
1384 |
1389 |
Electrical Requirements |
|
230V, 50Hz |
230V, 50Hz |
Dimensions |
Exterior Dimensions H x W x D inches (mm) |
61.8 x 51.2 x 31.5 (1568 x 1300 x 800) |
61.8 x 74.8 x 31.5 (1568 x 1900 x 800) |
Interior Dimensions H x W x D inches (mm)* |
30.7 x 47.2 x 24.8 (780 x 1200 x 630) |
30.7 x 70.9 x 24.8 (780 x 1800 x 630) |
Working Height of Front Window inches (mm) |
10 (254) |
10 (254) |
Maximum Opening Height of Front Window inches (mm) |
21 (535) |
21 (535) |
Shipping Dimensions H x W x D inches (mm) |
68.1 x 55.5 x 36.4 (1730 x 1410 x 925) |
68.1 x 79.1 x 36.4 (1730 x 2010 x 925) |
Weight |
Net Weight lbs (kg) |
375 (170) |
507 (230) |
Shipping Weight lbs (kg) |
430 (195) |
584 (265) |
Maximum Load of One-Piece Work Tray lbs (kg) |
110 (50) |
110 (50) |
Ventilation System |
Exhaust/Inflow Air Volume CFM (m3/h) |
342 (582) |
513 (872) |
Heat Emission |
Heat Emission at 25°C Ambient BTU/hr (kW) |
683 (0.2) |
1366 (0.4) |
Filter Specification |
Supply/Exhaust Air Filter |
H14 HEPA EN 1822 99.995% at the most penetrating particle size (mini-pleated) |
Performance |
Certification |
NSF/ANSI 49, UL |
NSF/ANSI 49, UL |
Sound Pressure Level dB (A) |
< 63 |
< 65 |
Lighting Power (fc) |
>120 |
>120 |
Electrical Data |
Power Consumption, Operating Set Point (kW) |
0.2 |
0.4 |
Current Consumption (Amps) |
1.2 |
1.9 |
Branch Circuit Protection |
T 15 Amp fuse or Class B 15 Amp circuit breaker is required. The local electrical regulations
in the country of use as well as the relevant connection conditions must be observed. |
Supply Management |
Receptacles |
The receptacles have a load capacity of up to 5 A and are protected with T 5 A fuses. When all
receptacles are in use simultaneously they must not exceed the maximum total load capacity of 5 A. |
Features |
Access Ports 0.94" (24 mm) diameter
Service Valves
Receptacles (Rear Wall) |
6 (3 on each side)
Up to 6 (installed through side wall access ports)
Two single, left and right side |