Operating Principle |
Combustion TOC(고온 연소산화 방식) High temperature (680 ˚C) catalytic combustion |
Wet Oxidation TOC(습식 산화 방식) Heated sodium persulfate oxidation |
Model(산화방식에 따른 모델 구분) |
1030C(Combustion) |
1030W(Wet Oxidation) |
1030D(Dual) Combustion and Wet Oxidation |
Measurement Technique |
Non-dispersive infrared (NDIR) detection |
Measurement Range |
100 ppb C – 30,000 ppm C |
10 ppb C – 30,000 ppm C |
Instrument Detection Limit (IDL) |
50 ppb C |
2 ppb C |
Operator Interface |
Color LCD touchscreen display with Windows® CE-based software |
Operating Modes |
Standalone (Windows® CE), PC-controlled, or LAN/ LIMS network connectivity |
Basic Software |
Single instrument operation with data transfer to PC |
Autosampler |
88 position rotary autosampler designed to fit directly underneath Aurora 1030D or 1030C/1030W analyzer |
Sample Injection |
Manual syringe, sipper tube, autosampler, or multi-stream at-line sampling module |
Certification |
CE, EMC: EN61326 / Safety: IEC 61010-11 2001 |
Reagents Required |
Hydrochloric acid, rinsewater |
Sodium persulfate, 5% phosphoric acid, rinsewater |
Sample Injection Volume |
10μL – 0.8 mL |
10μL – 10mL |
Method TC |
680 ˚C Pt catalyst, 900 ˚C non-catalyst packing |
Acid and persulfate reaction |
Method TIC |
Acidification and sparging |
Acidification with Phosphoric acid and sparging |
Method TOC |
NPOC, combustion of TIC-free sample, TOC.TIC |
NPOC by heated persulfate oxidation or TC-TIC |
Furnace Temperature |
Adjustable, 680 . 950 ˚C in 1 ˚C increments |
Adjustable to 100 ˚C in 1˚C increments |
Repeatability |
3.0% |
2.0% or 2 ppb |
Accuracy |
±2% FS or 2% relative, whichever is greater |
±1% FS or 2% relative, whichever is greater |
Sample pathway |
Color coded Teflon® tubing |
Sample handling |
Syringe with isolation loop to prevent contamination |
Gas Supply |
Zero-grade air, or O2 (99.998%) |
N2 (99.998%), zero-grade air, or O2 (99.998%) |
Power Supply |
Variable voltage, 100-240VAC, 50/60 Hz, 950W |