Metrological parameters |
Maximum capacity [Max] |
6 g |
Minimum load |
0,1 mg |
Readability [d] |
1 µg |
Verification unit [e] |
1 mg |
Tare range |
-6 g |
Standard repeatability [5% Max] |
1,3 µg |
Standard repeatability [Max] |
3,5 µg |
Standard minimum weight (USP) |
2,6 mg |
Standard minimum weight (U=1%, k=2) |
0,26 mg |
Permissible repeatability [5% Max] |
2 µg |
Permissible repeatability [Max] |
5 µg |
Linearity |
±9 µg |
Eccentric load deviation |
7 µg |
Sensitivity time drift |
1×10⁻⁶/Year×Rt |
Stabilization time |
~3,5 s |
Adjustment |
internal (automatic) |
OIML Class |
I |
Physical parameters |
Leveling system |
automatic - Reflex Level System |
Display |
10″ touchscreen |
Protection class |
IP 43 |
Weighing chamber dimensions |
199×170×217 mm |
Weighing pan dimensions |
ø30 mm + 2× intended for stents |
Packaging dimensions |
750×492×595 mm |
Net weight |
14.5 kg |
Gross weight |
18.9 kg |
Communication interface |
Communication interface |
USB-A ×2, USB-C, HDMI, Ethernet, Wi-Fi, Hotspot |
Electrical parameters |
Power supply |
Adapter: 100 – 240V AC 50/60Hz 1A; 15V DC 2.4A , Balance: 12 – 15V DC 1.4A max* |
Environmental conditions |
Operating temperature |
+10 — +40 °C |
Operating temperature change rate |
±0.3°C/1h (±1°C/8h) |
Relative humidity |
40% — 80% |
Relative humidity change rate |
±1%/h (±4%/8h) |