Metrological parameters |
| Maximum capacity [Max] |
6 g |
| Minimum load |
0,1 mg |
| Readability [d] |
1 µg |
| Verification unit [e] |
1 mg |
| Tare range |
-6 g |
| Standard repeatability [5% Max] |
1,3 µg |
| Standard repeatability [Max] |
3,5 µg |
| Standard minimum weight (USP) |
2,6 mg |
| Standard minimum weight (U=1%, k=2) |
0,26 mg |
| Permissible repeatability [5% Max] |
2 µg |
| Permissible repeatability [Max] |
5 µg |
| Linearity |
±9 µg |
| Eccentric load deviation |
7 µg |
| Sensitivity time drift |
1×10⁻⁶/Year×Rt |
| Stabilization time |
~3,5 s |
| Adjustment |
internal (automatic) |
| OIML Class |
I |
Physical parameters |
| Leveling system |
automatic - Reflex Level System |
| Display |
10″ touchscreen |
| Protection class |
IP 43 |
| Weighing chamber dimensions |
199×170×217 mm |
| Weighing pan dimensions |
ø30 mm + 2× intended for stents |
| Packaging dimensions |
750×492×595 mm |
| Net weight |
14.5 kg |
| Gross weight |
18.9 kg |
Communication interface |
| Communication interface |
USB-A ×2, USB-C, HDMI, Ethernet, Wi-Fi, Hotspot |
Electrical parameters |
| Power supply |
Adapter: 100 – 240V AC 50/60Hz 1A; 15V DC 2.4A , Balance: 12 – 15V DC 1.4A max* |
Environmental conditions |
| Operating temperature |
+10 — +40 °C |
| Operating temperature change rate |
±0.3°C/1h (±1°C/8h) |
| Relative humidity |
40% — 80% |
| Relative humidity change rate |
±1%/h (±4%/8h) |