Metrological parameters |
Maximum capacity [Max] |
21 g |
Minimum load |
0,1 mg |
Readability [d] |
1 µg |
Verification unit [e] |
1 mg |
Tare range |
-21 g |
Standard repeatability [5% Max] |
1 µg |
Standard repeatability [Max] |
3 µg |
Standard minimum weight (USP) |
2 mg |
Standard minimum weight (U=1%, k=2) |
0,2 mg |
Permissible repeatability [5% Max] |
1,6 µg |
Permissible repeatability [Max] |
4 µg |
Linearity |
±7 µg |
Eccentric load deviation |
7 µg |
Sensitivity time drift |
1×10⁻⁶/Year×Rt |
Stabilization time |
~5 s |
Adjustment |
internal (automatic) |
OIML Class |
I |
Physical parameters |
Leveling system |
automatic - Reflex Level System |
Display |
10″ touchscreen |
Delivery components |
Microbalance, terminal, weighing pan, weighing pan shield, glass
vessel, evaporation ring, glass lid, additional glass lid, glass lid, power supply, pincette, brush, fabric dust cover. |
Weighing chamber dimensions |
ø90×90 mm |
Weighing pan dimensions |
ø26 mm |
Packaging dimensions |
750×492×595 mm |
Net weight |
9,1 kg |
Gross weight |
16,6 kg |
Communication interface |
Communication interface |
USB-A ×2, USB-C, HDMI, Ethernet, Wi-Fi, Hotspot |
Electrical parameters |
Power supply |
Adapter: 100 – 240V AC 50/60Hz 1A; 15V DC 2,4A Balance: 12 – 15V DC 1,4A max* |
Environmental conditions |
Operating temperature |
+10 — +40 °C |
Operating temperature change rate |
±0,3°C/1h (±1°C/8h) |
Relative humidity |
40% — 80% |
Relative humidity change rate |
±1%/h (±4%/8h) |