| Select the REFRIGERATION TEMPERATURE of your
          freeze dryer’s collector. This is the second digit of your
          catalog number.
 0 = -50° C (-58° F). For samples containing water.
          Not for use with samples
 containing acetonitrile,
          methanol or ethanol. Collector is capable of removing
 2 liters of water in 24 hours.*
 1 = -84° C (-119° F). For low eutectic point samples
          containing water or acetonitrile. Not for use with
          samples containing methanol or ethanol. Collector is
          capable of removing 4 liters of water in 24 hours.*
 2 = -105° C (-157° F) at 60 Hz or -102° C (-152° F) at 50 Hz.
          For low eutectic point samples containing water,
          acetonitrile or dilute methanol or ethanol. Collector is
 capable of removing 2.5 liters of water in 24 hours.*
 | Select the COLLECTOR COIL and CHAMBER
          MATERIAL of your freeze dryer. This is the sixth digit
          of your catalog number.1 = Stainless steel
 2 = PTFE-coated stainless steel
 Select the ELECTRICAL CONFIGURATION and
          PLUG TYPE of your freeze dryer. These are the eighth
          and ninth digits of your catalog number.
 00 = 115V, 60 Hz, with 20A plug for North America
 10 = 230V, 60 Hz, with plug for North America
 15 = 230V, 60 Hz, with plug for Saudi Arabia
 30 = 230V, 50 Hz, with Schuko plug
 40 = 230V, 50 Hz, with plug for China/Australia
 50 = 230V, 50 Hz, with plug for Great Britain (UK)
 70 = 230V, 50 Hz, with plug for India
 |