| Specifications and Ordering Information | 4 ft models | 6 ft models | 
  
    | Standard Model) | Coated interior walls, cabinet only | 1354 | 1359 | 
  
    | Standard Package | Includes coated interior cabinet walls, manualadjustable height stand, factory installed UV light,
 and one set of armrests
 | 1384 | 1389 | 
  
    | Electrical Requirements |   | 230V, 50Hz | 230V, 50Hz | 
  
    | Dimensions | Exterior Dimensions H x W x D inches (mm) | 61.8 x 51.2 x 31.5 (1568 x 1300 x 800) | 61.8 x 74.8 x 31.5 (1568 x 1900 x 800) | 
  
    | Interior Dimensions H x W x D inches (mm)* | 30.7 x 47.2 x 24.8 (780 x 1200 x 630) | 30.7 x 70.9 x 24.8 (780 x 1800 x 630) | 
  
    | Working Height of Front Window inches (mm) | 10 (254) | 10 (254) | 
  
    | Maximum Opening Height of Front Window inches (mm) | 21 (535) | 21 (535) | 
  
    | Shipping Dimensions H x W x D inches (mm) | 68.1 x 55.5 x 36.4 (1730 x 1410 x 925) | 68.1 x 79.1 x 36.4 (1730 x 2010 x 925) | 
  
    | Weight | Net Weight lbs (kg) | 375 (170) | 507 (230) | 
  
    | Shipping Weight lbs (kg) | 430 (195) | 584 (265) | 
  
    | Maximum Load of One-Piece Work Tray lbs (kg) | 110 (50) | 110 (50) | 
  
    | Ventilation System | Exhaust/Inflow Air Volume CFM (m3/h) | 342 (582) | 513 (872) | 
  
    | Heat Emission | Heat Emission at 25°C Ambient BTU/hr (kW) | 683 (0.2) | 1366 (0.4) | 
  
    | Filter Specification | Supply/Exhaust Air Filter | H14 HEPA EN 1822 99.995% at the most penetrating particle size (mini-pleated) | 
  
    | Performance | Certification | NSF/ANSI 49, UL | NSF/ANSI 49, UL | 
  
    | Sound Pressure Level dB (A) | < 63 | < 65 | 
  
    | Lighting Power (fc) | >120 | >120 | 
  
    | Electrical Data | Power Consumption, Operating Set Point (kW) | 0.2 | 0.4 | 
  
    | Current Consumption (Amps) | 1.2 | 1.9 | 
  
    | Branch Circuit Protection | T 15 Amp fuse or Class B 15 Amp circuit breaker is required. The local electrical regulations
      in the country of use as well as the relevant connection conditions must be observed. | 
  
    | Supply Management | Receptacles | The receptacles have a load capacity of up to 5 A and are protected with T 5 A fuses. When all
      receptacles are in use simultaneously they must not exceed the maximum total load capacity of 5 A. | 
  
    | Features | Access Ports 0.94" (24 mm) diameterService Valves
 Receptacles (Rear Wall)
 | 6 (3 on each side)
      Up to 6 (installed through side wall access ports) Two single, left and right side
 |